Defibrillator end of life: recycling and operator obligations

Written by Cyrille GAGNAIRE, France Défibrillateur. Published on 12 September 2026.

An automated external defibrillator (AED) does not last forever. Its case, electrodes and battery eventually reach the end of their life, replaced or permanently taken out of service. Getting rid of these parts is not trivial: they contain electronic components and a lithium battery subject to specific collection channels. Here is how to properly retire a decommissioned unit, where to drop off each part, and the paperwork not to forget for the declaration and the safety logbook.

Why a defibrillator eventually gets taken out of service

A well maintained AED (defibrillator, DAE in French) works for eight to ten years, sometimes longer. Its two consumables age faster, though: electrodes expire after two to five years depending on the model, and the battery wears out after three to five years. Past these deadlines, or as soon as a self test reports a persistent fault, the unit must be pulled from service rather than left in place under a cloud of doubt about whether it still works.

Obsolescence can also come from the manufacturer: some older models no longer receive updates and spare parts are no longer available. In that case, replacing the whole unit is a better option than artificially extending its life with incompatible or dubious second hand consumables.

A hard knock, a fall from the wall cabinet or water damage can also end a unit's life prematurely, regardless of its age. The self test does not always catch this kind of mechanical damage, which is why a regular visual check of the case matters, in addition to reading the indicator lights.

What never to do with a decommissioned unit

A retired defibrillator must never stay displayed in its cabinet as if it were operational: someone on site could grab it in an emergency believing it works. It must be removed from its mount as soon as it is taken out of service, or clearly marked as unavailable while the replacement is pending.

Throwing the unit, the electrodes or the battery in with household waste is also to be avoided. The case contains electronic components, and the battery, most often lithium based, poses a fire risk if crushed or punctured in a regular waste collection truck.

Another common mistake is keeping a retired unit in a cupboard, unlabelled, with the idea of using it as a backup one day. A decommissioned AED left lying around is almost always forgotten, so it is better to arrange for its removal as soon as the replacement is in place.

Where to drop off each part

Each component follows a different channel. The electronic case falls under waste electrical and electronic equipment: it can be dropped at a recycling centre or taken back directly by the provider installing the new unit. The battery must be separated and taken to a dedicated collection point for batteries and accumulators, at a recycling centre or with certain electronics retailers, because of the risk associated with lithium. Used electrodes, once their gel has dried out and they carry no electrical risk, are generally treated as ordinary packaging waste, unless the manufacturer's instructions state otherwise.

Part removedSuitable recycling channel
AED caseWaste electrical and electronic equipment, at a recycling centre or taken back by the provider during replacement
BatteryCollection point for batteries and accumulators, at a recycling centre or with certain retailers
Used electrodesOrdinary packaging waste once the gel has dried, unless the manufacturer states otherwise

Updating the declaration and the logbook

A defibrillator that is removed, replaced or moved must be reported in Geo'DAE, the national registry of defibrillators, where declaration is mandatory. An entry pointing to a unit that no longer exists could mislead a bystander or emergency responders looking for the nearest AED.

The site's safety logbook must also record the removal date, the reason, and the commissioning date of the new unit where applicable. This traceability is part of the operator's maintenance obligations under article R. 5212-25 of the French public health code (Code de la santé publique).

This update only takes a few minutes, but it prevents a lasting gap between what a bystander or a locator app shows and the actual situation on site. An operator managing several sites should handle this at the same time as the physical replacement, so it does not get forgotten afterwards.

Planning the replacement instead of coping with a breakdown

Replacement should be planned before the unit actually fails. Tracking self test alerts and the expiry dates printed on the electrodes and battery makes it possible to order the replacement in time, without leaving the site unprotected.

To avoid the upfront cost of a full purchase, rental, starting at 50 euros excl. VAT per month with maintenance included, covers exactly this renewal of consumables and the replacement of the unit at end of life. For a purchase, the warranty runs for five years and delivery takes two to five business days, with a quote provided within 24 hours.

Frequently asked questions

Can a defibrillator that no longer works be thrown in the bin?

No. The electronic case falls under waste electrical and electronic equipment, and the lithium battery must be taken to a dedicated collection point for batteries and accumulators. Throwing them out with household waste creates a fire risk during collection and prevents these materials from being properly recycled.

Does removing a defibrillator need to be declared in Geo'DAE?

Yes. The Geo'DAE declaration must reflect the actual situation on site. A unit that is removed, replaced or moved must be updated in the registry so that a bystander or emergency responders are not directed to a defibrillator that no longer exists, or no longer sits at that location.

Can the provider take back the old unit when replacing it?

This is the simplest option in practice. Many providers take back the old unit and its used consumables when installing the new defibrillator, which guarantees a suitable recycling channel without extra steps for the operator. It is worth asking about this when placing the order, even before the new unit is delivered.

Is a defibrillator battery a particular hazard?

Yes, if it is lithium based. Punctured, crushed or exposed to strong heat, it can catch fire. That is why it must be kept separate from other waste and taken to a collection point specific to batteries and accumulators, at a recycling centre or with certain retailers.

What should be done with expired electrodes?

Expired electrodes should no longer stay in the defibrillator, since their dried conductive gel compromises contact during a shock. Once removed, they are generally treated as ordinary packaging waste, unless the manufacturer states otherwise in the product instructions.