Spare Pads: Should You Keep a Stock in Advance?
Written by Cyrille GAGNAIRE, France Défibrillateur. Published on 11 November 2026.

A defibrillator with expired pads or a weak battery no longer does its job, even if the casing looks perfect. Many operators order consumables at the last minute, sometimes after the self-test alert. Keeping a spare set avoids a period without protection, but poorly managed, the stock becomes a pointless expense. Here is how to decide, store and renew.
The real risk: a period without a usable device
When the self-test reports expired pads or a weak battery, the device is no longer reliable. If the person in charge then has to place an order, wait for delivery and fit the parts, the site has no working defibrillator for several days.
Yet in France around 50,000 cardiac arrests occur every year, and each minute without defibrillation reduces the chances of survival by about 10%. A few days of unavailability is therefore not an administrative detail: it directly exposes the people on site.
Maintenance of the defibrillator remains the operator's responsibility under article R. 5212-25 of the Code de la santé publique (French Public Health Code). Anticipating the replacement of consumables is part of that responsibility.
What is worth keeping in reserve
Not all parts are equal when it comes to stock. Pads are the priority, because they are single use: after any use, even a brief one, they must be replaced. The battery is less suited to stockpiling, because it loses capacity over time even when unused.
The table below gives general benchmarks, to be checked against your model's manual.
| Item | Indicative lifespan | Value of a stock |
|---|---|---|
| Adult pads | 2 to 5 years | High: single use, to be replaced after every intervention |
| Paediatric pads | 2 to 5 years | High if children visit the site |
| Battery | 3 to 5 years | Moderate: it also ages on the shelf |
| Device | 8 to 10 years and more | None: replacement is planned |

Storing without wasting
A spare pair of pads has its own expiry date, printed on the packaging. A set bought too early may expire almost at the same time as the one already fitted. When ordering, check the stated expiry date and favour parts with a long remaining life.
Keep consumables in their sealed original packaging, away from heat, frost and humidity, and out of direct light. An unheated technical room or a car boot in the sun are poor choices.
Store the spare set in a place known to the person in charge, but separate from the defibrillator cabinet: this avoids mixing up parts in an emergency and exposing them to the same temperature swings as the device.
Organising the renewal
A stock is only valuable if it is tracked. Record in the safety register the expiry date of the fitted pads, of the battery and of the spare set. A reminder set three months before the first deadline leaves time to order without pressure.
Appoint a responsible person, with a deputy. Without a clearly identified lead, the date goes unnoticed and nobody realises the spare set has aged too.
After any use of the defibrillator, immediately replace the pads with the reserve set, then buy a new spare. The device is available again the same day instead of waiting for a delivery, which generally takes 2 to 5 working days.
On-site stock or outsourced maintenance
Not every site needs a stock. A place with a single device, an available lead and a responsive supplier can simply order with a three-month margin. An isolated site, an establishment open all year or several devices in one place justify a spare set.
Rental with maintenance included, available from €50 excl. VAT per month, shifts part of this concern to the provider, who tracks consumable deadlines. It does not remove the need to check the device, but it reduces the risk of oversight.
Whatever solution you choose, the criterion is the same: the day someone needs it, the device must be ready, with valid pads and a battery in good condition.
Frequently asked questions
Should you keep a spare battery?
It is not essential for most sites. The battery degrades over time even when stored, and its indicative lifespan is 3 to 5 years. It is better to track its replacement date and order it a few months before expiry, except for an isolated or highly exposed site.
Do spare pads have an expiry date?
Yes. Each pair has its own expiry date printed on the packaging, and the indicative lifespan is 2 to 5 years. Check this date on delivery and record it in the register so you do not discover too late that the stock has expired.
What should be done with expired pads?
They must no longer be fitted to the device. Their disposal falls under health-related waste rules or equipment collection channels. Your supplier can tell you the right procedure for your model.
Where should the spare set be stored?
In its sealed packaging, away from heat, frost and humidity, in a place known to the person in charge. Avoid an unheated room or a vehicle exposed to the sun, which damage consumables before they are even used.
Does a maintenance contract replace the stock?
For many sites it can, since the provider tracks deadlines and supplies the parts. Depending on the contract, check who delivers pads after use and how quickly. Without a precise commitment, a spare set remains prudent.
Planning an AED project?
Purchase, rental from 50 EUR excl. VAT per month, installation with GéoDAE registration, servicing and training: our team replies within 24 hours.